3.9 Status Bar
At the bottom of the Manual screen is a status bar that is continually updated
with system parameters.
Section 4. Anion Exchange Separation of Protein Standards
The starter kit enables you to learn to use the DuoFlow Maximizer or Pathfinder
systems by programming and running separation of a premixed anion exchange
standard containing equine myoglobin, conalbumin, chicken ovalbumin, and
soybean trypsin inhibitor, using a 1.3 ml UNO Q1 column (catalog #720-0001).
Equine myoglobin is not retained on the UNO Q1 column and elutes in the void
volume. Conalbumin, chicken ovalbumin, and soybean trypsin inhibitor bind to
the column and require increased salt concentrations for elution. Separation
requires approximately 6 minutes.
4.1 Overview of the Procedure
Run Conditions
•
Buffer A1
50 mM Tris-HCl
•
Buffer A2
50 mM Tris base
•
Solution B1
deionized water
•
Solution B2
2.0 M NaCI
•
Flow rate
4.00 ml/min
•
Sample volume
50 µl
•
UV detection
0.1 AUFS
•
QuadTec detection
0.1 AUFS (
λ = 280 nm), 0.1 AUFS (λ = 260 nm),
1.0 AUFS (
λ = 214nm), and 0.4 AUFS (λ = 304 nm)
•
Conductivity 100
mS/cm
•
pH
pH 7.1 to pH 9.1
General Procedure
Step 1
Prepare solutions A
1
, A
2
, B
1
, and B
2
Step 2
Prepare sample
Step 3
Install the UNO Q1 column
4006208B.qxd 6/28/2004 11:03 AM Page 29
- BioLogic DuoFlow Pathfinder 20 System DuoFlow Chromatography System Starter Kit BioLogic DuoFlow Maximizer 80 System DuoFlow Chromatography System Starter Kit BioLogic DuoFlow Maximizer 20 System DuoFlow Chromatography System Starter Kit BioLogic DuoFlow QuadTec 40 System DuoFlow Chromatography System Starter Kit BioLogic DuoFlow QuadTec 10 System DuoFlow Chromatography System Starter Kit BioLogic DuoFlow 40 System DuoFlow Chromatography System Starter Kit BioLogic DuoFlow 10 System DuoFlow Chromatography System Starter Kit