beautypg.com

Bio-Rad Trans-Blot® Plus Cell User Manual

Page 16

background image

Table 3.3 Isoelectric Focusing, Native, Acid Urea Gels, Basic
Proteins

These conditions were determined empirically using 12.5% Tris-HCl Criterion

gels and native horse myoglobin samples. Read about acidic transfers in Section 2.7
of this manual.

High Intensity Field Conditions (Cooling Required)

PLATE ELECTRODE DISTANCE

4 cm

7 cm

10 cm

Buffer Power

Run

Power

Run

Power

Run

conditions

Time conditions Time conditions Time

0.7% Acetic

50V/2A

45 min

70V/2A

60 min

100V/1.9A

60 min

acid , pH 2.8

70V/3A

15–30 min

110V/3A

15–30 min

125V/2.4A 15–30 min

Standard Field Conditions (Cooling Recommended)

PLATE ELECTRODE DISTANCE

4 cm

7 cm

10 cm

Buffer Power

Run

Power

Run

Power

Run

conditions Time conditions Time conditions Time

0.7% Acetic

10V/0.4A

16 Hrs.

10V/0.25A

16 Hrs.

10V/0.15A

16 Hrs.

acid , pH 2.8

20V/0.8A

20V/0.55A

20V/0.35A

30V/0.8A

30V/0.55A

14