Table 3.1 sds-page gels – Bio-Rad Trans-Blot® Plus Cell User Manual
Page 14
Table 3.1 SDS-PAGE Gels
These conditions were determined empirically using 12.5% Tris-HCl Criterion gels
and prestained SDS-PAGE molecular weight standards (Catalog # 161-0318). See
Section 3.3 for buffer formulations.
High Intensity Field Conditions (Cooling Required)
PLATE ELECTRODE DISTANCE
4 cm
7 cm
10 cm
Buffer Power
Run
Power
Run
Power
Run
conditions
Time conditions
Time
conditions
Time
Towbin
60V/2A*
45 min
100V/2A
45 min
100V/1.5A
60 min
90V/3A
15–30 min
120V/2.4A 15–30 min
130V/~2.3A 15–30 min
CAPS
41V/2A
15 min
66V/2A
30 min
95V/2A
30 min
60V/3A
15 min
100V/3A
15–30 min
120V/2.5A 15–30 min
Carbonate
25V/2A
15 min
40V/2A
30 min
55V/2A
30 min
35V/3A
15 min
55V/3A
15–30 min
80V/3A
15–30 min
Standard Field Conditions (Cooling Recommended)
PLATE ELECTRODE DISTANCE
4 cm
7 cm
10 cm
Buffer Power
Run
Power
Run
Power
Run
conditions Time conditions
Time conditions Time
Towbin
10V/0.3A
16 Hrs.
10V/0.2A
16 Hrs.
10V/0.15A
16 Hrs.
20V/ 0.7A
20V/0.4A
20V/0.3A
30V/ 1.1A
30V/0.7 A
30V/0.5 A
CAPS
10V/ 0.35A
4 Hrs.
10V/0.2A
4 Hrs.
10V/0.2A
4 Hrs.
20V/0.8A
20V/0.55 A
20V/0.4A
30V/0.9A
30V/0.6A
Carbonate
10V/0.7A
4 Hrs
10V/0.4A
4 Hrs
10V/0.33A
4 Hrs
20V/1A
20V/0.7A
30V/1A
* When running under constant voltage, the PowerPac 200 power supply will automatically cross over to constant
current if the power supply's current limit is reached.
12