beautypg.com

Table 3.1 sds-page gels – Bio-Rad Trans-Blot® Plus Cell User Manual

Page 14

background image

Table 3.1 SDS-PAGE Gels

These conditions were determined empirically using 12.5% Tris-HCl Criterion gels

and prestained SDS-PAGE molecular weight standards (Catalog # 161-0318). See
Section 3.3 for buffer formulations.

High Intensity Field Conditions (Cooling Required)

PLATE ELECTRODE DISTANCE

4 cm

7 cm

10 cm

Buffer Power

Run

Power

Run

Power

Run

conditions

Time conditions

Time

conditions

Time

Towbin

60V/2A*

45 min

100V/2A

45 min

100V/1.5A

60 min

90V/3A

15–30 min

120V/2.4A 15–30 min

130V/~2.3A 15–30 min

CAPS

41V/2A

15 min

66V/2A

30 min

95V/2A

30 min

60V/3A

15 min

100V/3A

15–30 min

120V/2.5A 15–30 min

Carbonate

25V/2A

15 min

40V/2A

30 min

55V/2A

30 min

35V/3A

15 min

55V/3A

15–30 min

80V/3A

15–30 min

Standard Field Conditions (Cooling Recommended)

PLATE ELECTRODE DISTANCE

4 cm

7 cm

10 cm

Buffer Power

Run

Power

Run

Power

Run

conditions Time conditions

Time conditions Time

Towbin

10V/0.3A

16 Hrs.

10V/0.2A

16 Hrs.

10V/0.15A

16 Hrs.

20V/ 0.7A

20V/0.4A

20V/0.3A

30V/ 1.1A

30V/0.7 A

30V/0.5 A

CAPS

10V/ 0.35A

4 Hrs.

10V/0.2A

4 Hrs.

10V/0.2A

4 Hrs.

20V/0.8A

20V/0.55 A

20V/0.4A

30V/0.9A

30V/0.6A

Carbonate

10V/0.7A

4 Hrs

10V/0.4A

4 Hrs

10V/0.33A

4 Hrs

20V/1A

20V/0.7A

30V/1A

* When running under constant voltage, the PowerPac 200 power supply will automatically cross over to constant

current if the power supply's current limit is reached.

12