beautypg.com

Bio-Rad Trans-Blot® Plus Cell User Manual

Page 15

background image

Table 3.2 Native Gels

These conditions were determined empirically using 12.5% Tris-HCl Criterion

gels and native horse myoglobin samples. See Section 3.3 for buffer formulations.

High Intensity Field Conditions (Cooling Required)

PLATE ELECTRODE DISTANCE

4 cm

7 cm

10 cm

Buffer Power

Run

Power

Run

Power

Run

conditions Time conditions Time conditions Time

Towbin

50V/2A

45 min

90V/2A

45 min

100V/1.7A

60 min

(no methanol)

80V/3A

15–30 min

100V/2.2A 15–30 min

130V/2.3A 15–30 min

Standard Field Conditions (Cooling Recommended)

PLATE ELECTRODE DISTANCE

4 cm

7 cm

10 cm

Buffer Power

Run

Power

Run

Power

Run

conditions Time conditions Time conditions Time

Towbin

10V/0.3A

16 Hrs.

10V/0.2A

16 Hrs.

10V/0.15A

16 Hrs.

(no methanol)

20V/0.7A

20V/0.4A

20V/0.3A

30V/1.1A

30V/0.7A

30V/0.5A

13