Metrohm 795 KFT Titrino User Manual
Page 91
3.2 Remote control commands
795 KFT Titrino
87
.CyclNo: Cycle number. Together with the cycle time (see 3.2.2.75), a time
frame can be set up. The other data belong to the corresponding
cycle number. The cycle number is set to 0 at the start of a
method and it is incremented until the end of the method.
.V: Volume.
.dVdt:
associated volume drift.
.dMeasdt: associated measured value drift.
.dMeasdV: associated 1st derivative of the titration curve.
The unit "titrator" must be preset (see 3.2.2.90).
3.2.2.92. Setup.AutoInfo.Status ON, OFF
Setup.AutoInfo.P ON, OFF
Setup.AutoInfo.T.R ON, OFF
Setup.AutoInfo.T.G ON, OFF
Setup.AutoInfo.T.GC ON, OFF
Setup.AutoInfo.T.S ON, OFF
Setup.AutoInfo.T.B ON, OFF
Setup.AutoInfo.T.F ON, OFF
Setup.AutoInfo.T.E ON, OFF
Setup.AutoInfo.T.H ON, OFF
Setup.AutoInfo.T.C ON, OFF
Setup.AutoInfo.T.O ON, OFF
Setup.AutoInfo.T.N ON, OFF
Setup.AutoInfo.T.Re ON, OFF
Setup.AutoInfo.T.Si ON, OFF
Setup.AutoInfo.T.M ON, OFF
Setup.AutoInfo.T.EP ON, OFF
Setup.AutoInfo.T.RC ON, OFF
Setup.AutoInfo.C.B1 ON, OFF
Setup.AutoInfo.C.R1 ON, OFF
Setup.AutoInfo.C.B2 ON, OFF
Setup.AutoInfo.C.R2 ON, OFF
Setup.AutoInfo.I ON, OFF
Setup.AutoInfo.O ON, OFF
ON means that the Titrino reports automatically the moment the correspond-
ing change occurs.
.Status:
Global switch for all set AutoInfo.
.P
PowerOn: Simulation of power on (3.2.2.94). Not from mains.
Messages from node .T, Titrator:
.T.R
Ready: Status 'Ready' has been reached.
.T.G
Go: Instrument has been started.
.T.GC
GoCommand: Instrument has received a go command.
.T.S
Stop: Status 'Stop' has been reached.
.T.B
Begin of sequence.
.T.F
Final: End of determination, the final steps will be carried out.
.T.E
Error. Message together with error number, see page 49ff.
.T.H
Hold: Status 'Hold' has been reached.
.T.C
Continue: Continue after hold.
.T.O
Conditioning OK: EP reached (in KFT with conditioning).
.T.N
Conditioning Not OK: EP not reached (in KFT with conditioning).
.T.Re
Request: In the inquiry of an identification or the sample size after
start of titration.