Metrohm 795 KFT Titrino User Manual
Page 76
3.2 Remote control commands
795 KFT Titrino
72
3.2.2.26. Mode.Parameter.Presel.IReq
id1, id1&2, all, OFF
Mode.Parameter.Presel.SReq
value, unit, all, OFF
Automatic inquiry after the start of the determination. From such an inquiry,
the determination continues if the requested entry/entries is/are made, e.g.
&SmplData.OFFSilo.Id1 (see 3.2.2.52) or with &M $G, see 3.2.2.1.
$H is not possible in requests.
3.2.2.27. Mode.Parameter.Presel.LimSmplSize.Status ON,
OFF
Mode.Parameter.Presel.LimSmplSize.LoLim
0.0...999 999
Mode.Parameter.Presel.LimSmplSize.UpLim 0.0...999 999
Limit control for the sample size.
3.2.2.28. Mode.Parameter.Presel.Oven COM1,
COM2,
no
If an oven is connected, its result will be incorporated into the result report of
the Titrino. If there is no oven connected via RS232, the setting of this pa-
rameter has to be "no".
3.2.2.29. Mode.Parameter.Presel.ActPuls
first, all, cond., OFF
Output of a pulse on the I/O line "Activate", see page 126.
3.2.2.30. UserMeth.FreeMem read
only
Memory space, available for user methods or silo lines. $Q sends the number
of free bytes, e.g. "4928".
3.2.2.31. UserMeth.Recall $G
UserMeth.Recall.Name
up to 8 ASCII characters
UserMeth.Store $G
UserMeth.Store.Name
up to 8 ASCII characters
UserMeth.Delete $G
UserMeth.Delete.Name
up to 8 ASCII characters
UserMeth.DelAll $G
Management of the internal method memory: Load, store and delete methods.
An action is performed if "$G" is sent to the corresponding node just after en-
tering the name.
Do not use blank characters before and after method name!
.DelAll: Deletes all methods in the user memory.