3.2 Remote control commands
795 KFT Titrino
81
3.2.2.68. Info.ActualInfo.Assembly.Counter.V read
only
Info.ActualInfo.Assembly.Counter.Clear $G
$Q sends the volume. With the function &Info.Assembly.Counter.Clear $G, the
volume counter is set to zero.
3.2.2.69. Info.ActualInfo.Assembly.Meas read
only
$Q sends the current measured value from the assembly.
3.2.2.70. Info.ActualInfo.Titrator.CyclNo read
only
Info.ActualInfo.Titrator.V read
only
Info.ActualInfo.Titrator.Meas read
only
Info.ActualInfo.Titrator.dVdt read
only
Info.ActualInfo.Titrator.dMeasdt read
only
Info.ActualInfo.Titrator.dMeasdV read
only
Info.ActualInfo.Titrator.T read
only
$Q sends the current values in the following formats:
KFT
CyclNo 127
V(mL) 1.2345
Meas:
Ipol (mV)
Upol (µA)
–345.6
–12.5
dVdt (µL/s)
2.5142
dMeasdt
Ipol (mV/s)
Upol (µA/s)
0.7957
0.7957
dMeasdV (mV/µL)
10.6326
NV: Not Valid.
OV will be sent for "overrange".
A time frame can be set up from the cycle number and the cycle time (see
3.2.2.75). The cycle number is set to 0 at the start of a method and it is in-
cremented until the end of the method.
3.2.2.71. Info.ActualInfo.MeasPt.Index read
only
Info.ActualInfo.MeasPt.X read
only
Info.ActualInfo.MeasPt.Y read
only
Info.ActualInfo.MeasPt.Z1 read
only
Info.ActualInfo.MeasPt.Z2 read
only
Info.ActualInfo.EP.Index read
only
Info.ActualInfo.EP.X read
only
Info.ActualInfo.EP.Y read
only
$Q sends the last entry into the measuring point list (.MeasPt) or the last entry
into the list of EP's.