3.2 Remote control commands
795 KFT Titrino
78
Reports which are sent from the Titrino are marked with space (ASCII 32) and
' at the beginning. Then an individual identifier for each report follows. Reports
which are triggered by RS232 ($G) have the same introducer but without pre-
ceding space, i.e. they start with '.
3.2.2.61. Info.Checksums $G
Info.Checksums.MPList read
only
Info.Checksums.ActualMethod read
only
The checksums can be used to identify the content of a file unequivocally,
e.g. files with identical content
have identical results of the checksums. An empty file has checksum "0". The
calculation of the checksums is triggered with $G.
.MPList: Result of the checksum of the current measuring point list.
.ActualMethod: Result of the checksum of the current method in the working
memory. Identical methods with different method names have the
same results of the checksum.
3.2.2.62. Info.DetermData $G
Info.DetermData.Write ON,
OFF
Info.DetermData.MPList.1.Attribute read
only/read+write
Info.DetermData.ExV read
only/read+write
Info.DetermData.MPList.1.X read
only/read+write
Info.DetermData.MPList.1.Y read
only/read+write
Info.DetermData.MPList.1.Z1 read
only/read+write
Info.DetermData.MPList.1.Z2 read
only/read+write
for every measuring point
Determination data in hexadecimal format. A measuring point list is available.
Recalculation of the measuring data is triggered with $G.
.Write:
With "ON", the following nodes can be overwritten:
&Info.DetermData.MP.List, &Info.TitrResults.Var.C4X (X = 0...5),
and &Mode.Name.
.ExV:
Volume of the exchange unit, with which the determination was
executed
.MPList.1.Attribute: Attribute
.MPList.X: X coordinate, time
.MPList.Y: Y coordinate, volume
.MPList.Z1: Z1 coordinate, measuring value
.MPList.Z2: without meaning