Metrohm 795 KFT Titrino User Manual
Page 75
3.2 Remote control commands
795 KFT Titrino
71
3.2.2.22. Mode.Parameter.StopCond.VStop.Type abs., rel., OFF
Mode.Parameter.StopCond.VStop.V 0...99.99...9999.99
Mode.Parameter.StopCond.VStop.Factor 0...±999999
Stop volume.
If an absolute stop volume (abs.) has been selected, the volume in mL is
valid.
A relative stop volume (rel.) is dispensed as a function of the sample size:
Stop volume in mL = smpl size ∗ factor
The factor is valid.
OFF means that the criterion is not monitored.
3.2.2.23. Mode.Parameter.StopCond.FillRate 0.01...150, max.
Filling rate in the titration in mL/min. Max. means maximum possible filling
rate with the Exchange Unit in current use.
3.2.2.24. Mode.Parameter.Statistics.Status ON, OFF
Mode.Parameter.Statistics.MeanN
2...20
Mode.Parameter.Statistics.ResTab.Selected original, delete n,
delete
all
Mode.Parameter.Statistics.ResTab.DelN
1...20
Entries for the statistics calculations.
.Status:
On/off switching. Requirement for statistics calculations is a valid
assignment, see 3.2.2.9.
.MeanN: Number of individual results for statistics calculations.
.ResTab.Select: Selection of the table for the statistics calculations.
original: Original table. The original table is (again) set up, i.e. any
individual results which have been deleted are reincorporated
in the statistics calculations.
delete n: Single result lines are removed from the statistics calcu-
lation. All results of the corresponding line in the statistics
table are deleted. Specification of the line number in
.ResTab.DelN.
delete all: Clear entire statistics table. The results can not be reac-
tivated.
.ResTab.DelN: Specification of the line number to be deleted.
3.2.2.25. Mode.Parameter.Presel.Cond
ON, OFF
Mode.Parameter.Presel.DriftDisp ON, OFF
Mode.Parameter.Presel.DCor.Type auto,
man., OFF
Mode.Parameter.Presel.DCor.Value 0.0...99.9
.Cond: Conditioning ON/OFF
.DriftDisp: Drift display during conditioning ON/OFF.
.DCor.Type: Type of drift take-over for the drift correction. auto: Take-over of
the drift value at start.
.DCor.Value: Drift value for the manual drift correction.