Mode – Metrohm 795 KFT Titrino User Manual
Page 57
3.2 Remote control commands
795 KFT Titrino
53
&Mode
Object Description
Input
range
Reference
& Root
├
Mode
Mode
$G, $S, $H, $C
3.2.2.1.
:
├
.
QuickMeas
Rapid meas. in basic mode
$G, $S
3.2.2.2.
├
.
Select
Mode selection
KFT 3.2.2.3.
├
.KFTQuantity
Measured quantity for KFT
Ipol, Upol
ditto
├
.Name
Name of current method
read only/read+write
3.2.2.4.
├
.Parameter*
Parameter of current mode, see below
│
├
.Def
Definitions for data output
│ ├
.Formulas
Calculation formulas
│ │
├
.1
for result 1
│ │
│
├
.Formula Calculation
formula
special
3.2.2.5.
│ │
│
├
.TextRS
Text for result output
up to 8 ASCII char
ditto
│ │
│
├
.Decimal
Number of decimal places
0...2...5 ditto
│ │
│
├
.Unit
Unit for result output
up to 6 ASCII char
ditto
│ │
│
├
.Limits
Limits for result
ON, OFF ditto
│ │
│
├
.LoLim Lower
limit
0...±999 999
ditto
│ │
│
├
.UpLim Upper
limit
0...±999 999
ditto
│ │
│
├
.Output
Output on L13
active, pulse, OFF ditto
│ │
│
:
up to 9 results
│ ├
.SiloCalc
Silo calculations
│ │
├
.Assign Assignment
│ │
│
├
.C24
Store as variable C24
RSX,EPX,CXX
3.2.2.6.
│ │
│
├
.C25
Store as variable C25
RSX,EPX,CXX
│ │
│
├
.MatchId
Matching of Id's
id1, id1&2, all, OFF
│ ├
.ComVar
Assignment of common variables
│ │
├
.C30 for
C30
RSX,EPX,CXX,MNX
3.2.2.7.
│ │
├
up to C39
│ ├
.Report
Reports at the end of determination
│ │
├
.Assign1
Output to COM1
special
3.2.2.8.
│ │
├
.Assign2
Output to COM 2
as COM1
│ ├
.Mean
Assignment for mean calculation
│ │
├
.1 MN1
│ │
│
├
.Assign
Input of variable
RSX, EPX, CXX
3.2.2.9.
│ │ │
:
│ ├
.TempVar without
meaning
├
.CFmla
Calculation constants
│
├
.1
Calculation constant C01
│ │
├
.Value
Input of value
0...±999999 3.2.2.10.
│
├
up to C19