Metrohm 795 KFT Titrino User Manual
Page 73
3.2 Remote control commands
795 KFT Titrino
69
3.2.2.11. Mode.Parameter.CtrlPara.EP Ipol:
0...250...±2000
Upol:
0...25...±200.0
Mode.Parameter.CtrlPara.UnitEp read
only
Setting of the EP, resp. Control point in mV (with Ipol) or µA (with Upol). The
corresponding unit can be read with .UnitEP.
3.2.2.12. Mode.Parameter.CtrlPara.Dyn Ipol: 1...100...2000
Upol:
0.1...10...200.0
Mode.Parameter.CtrlPara.UnitDyn read
only
Mode.Parameter.CtrlPara.MaxRate 0.01...10...150, max.
Mode.Parameter.CtrlPara.MinIncr 0.1...9.9, min.
Control parameters.
.Dyn: Dynamics (control range) in mV (with Ipol) or µA (with Upol). The cor-
responding unit can be read with .UnitDyn.
.MaxRate: Maximum allowed titration rate in mL/min. Max. means maximum
possible rate with mounted Exchange Unit.
.MinIncr: Minimum volume increment in µL.
OFF means that the criterion is not monitored.
3.2.2.13. Mode.Parameter.CtrlPara.Stop.Type drift, time
Mode.Parameter.CtrlPara.Stop.Drift 1...20...999
Mode.Parameter.CtrlPara.Stop.Time 0...10...999, inf.
Mode.Parameter.CtrlPara.Stop.StopT 0...99999, OFF
Type and size of the stop criterion of the titration.
.Type: Type of stop criterion after stop drift or switch-off delay time.
.Drift: Stop drift in µL/min. Applies when "drift" has been selected.
.Time: Switch-off delay time in s. Applies when "time" has been selected.
"inf." means infinite.
.StopT: Stop time in s. Applies when "time" has been selected and the value of
.Time is set to "inf.".
3.2.2.14. Mode.Parameter.TitrPara.Direction +,
–, auto
Titration direction.
"auto" means the titration direction is determined automatically by the instru-
ment.
3.2.2.15. Mode.Parameter.TitrPara.XPause
0...999999
Pause time in s. Runs before dosing the start volume.