3.2 Remote control commands
795 KFT Titrino
74
3.2.2.39. Config.PeriphUnit.Stirrer
ON, OFF
Automatic stirrer control. With "ON" the stirrer will be switched on after start-
ing a method. At the end of the method it is switched off again.
3.2.2.40. Config.PeriphUnit.RemoteBox.Status
ON, OFF
Config.PeriphUnit.RemoteBox.Keyboard US, deutsch, francais,
español,
schweiz.
Config.PeriphUnit.RemoteBox.Barcode input, method, id1, id2,
id3, smpl size
Connections via Remote Box.
.Status:
Select if a Remote Box is connected.
.Keyboard: Type of keyboard which is connected to the Remote Box.
.Barcode: Select target in Titrino where you wish to have the string from the
barcode reader. "input" means that the string comes into the field
where the cursor is currently placed.
3.2.2.41. Config.Aux.Language english, deutsch, francais, español,
italiano, portugese, svenska
Selection of the dialog language.
3.2.2.42. Config.Aux.Set $G
Config.Aux.Set.Date YYYY-MM-DD
Config.Aux.Set.Time hh:mm
Date and time.
Input format of the date: Year-month-day, two-digit, enter leading zeros.
Input format for the time: Hours:minutes, two-digit, enter leading zeros.
Date and time have to be set with &Config.Aux.Set $G just after entry of the
value.
3.2.2.43. Config.Aux.RunNo
0...9999
Current sample number.
Set to 0 on power on and initialization. After 9999, counting starts again at 0.
3.2.2.44. Config.Aux.AutoStart 1...9999, OFF
Number of automatic, internal starts.
3.2.2.45. Config.Aux.StartDelay
0...999999
Start delay time in s. During this time, the data of the preceding determination
are retained.
3.2.2.46. Config.Aux.ResDisplay bold, standard
Character set for the result display at the end of the determination.