Metrohm 795 KFT Titrino User Manual
Page 77
3.2 Remote control commands
795 KFT Titrino
73
3.2.2.32. UserMeth.List.1.Name read
only
UserMeth.List.1.Mode read
only
UserMeth.List.1.Quantity read
only
UserMeth.List.1.DosUnit read
only
UserMeth.List.1.Bytes read
only
UserMeth.List.1.Checksum read
only
for each method
List of the methods in the user method memory with the following character-
istics:
.Name:
Name of the method
.Mode: Mode
.Quantity: Measured
quantity
.DosUnit: Buret of the method
.Bytes:
Number of bytes of the user memory used by the method
.Checksum: Checksum of the method, see 3.2.2.61.
3.2.2.33. Config.Monitoring.Validation.Status ON,
OFF
Config.Monitoring.Validation.Interval 1...365...9999
Config.Monitoring.Validation.Counter
0...9999
Config.Monitoring.Validation.ClearCount $G
Monitoring of validation.
.Interval: Time interval in days for validation.
.Counter: Time counter in days since last validation.
.ClearCount: Clears the above counter.
3.2.2.34. Config.Monitoring.Service.Status ON,
OFF
Config.Monitoring.Service.Date YYYY-MM-DD
Monitoring of service interval.
3.2.2.35. Config.Monitoring.DiagRep ON,
OFF
Printing of system test report after each switching on of the Titrino.
3.2.2.36. Config.PeriphUnit.CharSet1
Epson, Seiko, Citizen, HP, IBM
Config.PeriphUnit.CharSet2
Selection of the character set and the graphics control characters for COM1
resp. COM2 of the Titrino.
IBM means the IBM character set following character set table 437 and IBM
graphics control characters. Select 'IBM' for work with the computer.
3.2.2.37. Config.PeriphUnit.RepToComport
1, 2, 1&2
Selection of COM of the Titrino where manually triggered reports should be
outputted.
3.2.2.38. Config.PeriphUnit.Balance Sartorius,Mettler,Mettler AT,
AND,Precisa
Selection of the balance type.