Config – Metrohm 795 KFT Titrino User Manual
Page 60
3.2 Remote control commands
795 KFT Titrino
56
&Config
Object Description
Input
range
Reference
& Root
:
├
Config
Instrument configuration
:
├
.Monitoring
Monitoring functions
│ ├
.Validation Validation
monitoring
3.2.2.33.
│ │ ├
.Status
Status of validation monitoring
ON, OFF
ditto
│ │ ├
.Interval
Time interval for validation
1...365...9999 ditto
│ │ ├
.Counter Time
counter
0...9999 ditto
│ │ ├
.ClearCount
Clears the counter above
$G
ditto
│ ├
.Service
Monitoring of Metrohm service
$G
3.2.2.34.
│ │ ├
.Status
Status of service monitoring
ON, OFF
ditto
│ │ ├
.Date
Date of next service
YYYY-MM-DD
ditto
│ ├
.DiagRep
Printing of system test report
ON, OFF 3.2.2.35.
│
├
.PeriphUnit
Selection of peripheral units
│ ├
.CharSet1 External
printer Epson,Seiko,Citizen
│ │
IBM,HP 3.2.2.36.
│ ├
.CharSet2
as for CharSet1
│ ├
.RepToComport Output of manual reports
1, 2, 1&2
3.2.2.37.
│ ├
.Balance
Selection of balance
Sartorius,Mettler,Mettler AT
│ │
AND,Precisa 3.2.2.38.
│ ├
.Stirrer Stirrer
control
ON,
OFF 3.2.2.39.
│ ├
.RemoteBox
Connected remote box
3.2.2.40.
│ │ ├
.Status Status
ON,
OFF ditto
│ │ ├
.Keyboard
Type of keyboard
US, deutsch, francais,
│ │ │
español,
schweiz.
ditto
│ │ ├
.Barcode
Input of barcode reader
input, method, id1, id2,
│ │ │
id3, smpl size
ditto
│
├
.Aux
Miscellaneous
│ ├
.Language Dialog
language
english, deutsch,
│ │
français, español, italiano,
│ │
portugese, svenska
3.2.2.41.
│ ├
.Set
Setting of date and time
$G
3.2.2.42.
│ │ ├
.Date Date
YYYY-MM-DD
│ │ ├
.Time Time
hh:mm
│ ├
.RunNo Run
number
0...9999 3.2.2.43.
│ ├
.AutoStart Automatic
start
1...9999,
OFF
3.2.2.44.
│ ├
.StartDelay
Start delay time
0...999999 3.2.2.45.
│ ├
.ResDisplay Result
display
standard,bold 3.2.2.46.
│ ├
.DevName
Device label
8 ASCII char.
3.2.2.47.
│ ├
.Prog
Program version
read only
3.2.2.48.