3.1 General rules
795 KFT Titrino
49
Status conditions of the global $S:
$S .Mode.KFT.QuickMeas:
Quick manual measurement from the initial status in mode KFT.
The instrument gives the status from which it has been stopped. The detailed
status information is therefore identical the information for the global status
$G.
Violation of monitored limits with action "end" give the status message
$S.Mode.KFT.Inac;EYYY.
3.1.4 Error messages
Error messages are added to the status messages and separated from them
by the sign ";".
E20
Check exchange unit.
Exit: Mount Exchange Unit (properly) or &m $S.
E21
Check electrode, short circuit.
Exit: Rectify fault or &m $S.
E22
Check electrode, break.
Exit: Rectify fault or &m $S.
E23
Division by zero.
Exit: The error message disappears on next startup or on recalculation.
E26
Manual stop.
Exit: The error message disappears on next startup.
E27
Stop V reached in KFT.
Exit: The error message disappears on next startup.
E28
Wrong object call up.
Exit: Send correct path for object. Start path at root.
E29
Wrong value or no value allowed.
Exit: Send correct value or call up new object.
E30
Wrong trigger, this trigger is not allowed or carrying-out of action not
possible.
Exit: Send correct trigger (exception: $D) or call up new object.
E31
Command is not possible in active status. Repeat command in inactive
status.
Exit: Send new command.
E32
Command is not possible during titration. Repeat command during the
conditioning phase or in inactive status.
Exit: Send new command.
E33
Value has been corrected automatically.
Exit: Send new command.
E34
Instrument at the end of the titration and sample data is edited; the instrument
at rest or editing during filling.
Exit: &m $S.