3.2 Remote control commands
795 KFT Titrino
79
3.2.2.63. Info.TitrResults.RS.1.Value read
only
etc., up to .9
Info.TitrResults.EP.1.V read
only
Info.TitrResults.EP.1.Meas read
only
etc., up to .2
Info.TitrResults.Var.C40 read
only/read+write
etc., up to .C45
Info.TitrResults.Var.DTime read
only/read+write
.RS: Values of the calculated results.
.EP: Endpoints:
Volume coordinate in mL, e.g. "1.2340"
Measured value coordinate in mV (with Ipol) "–241" or µA (with
Upol) "43.7".
.Var: Various variables. You may overwrite the variables C40...C45, see
3.2.2.62.
C40: Initial measured value in mV (with Ipol) "41" or µA (with
Upol).
C41: End volume in mL, "12.5360".
C42: Time from start of titration to end in s, "62".
C43: Volume drift on start of KFT titration from the conditioning in
µL/min, "3.5".
C44: Temperature in °C.
C45: Start volume in mL, "2.800".
DTime: Time for the drift correction in KFT with conditioning.
3.2.2.64. Info.StatisticsVal.ActN read
only
Info.Statistics.1.Mean read
only
Info.Statistics.1.Std read
only
Info.Statistics.1.RelStd read
only
etc. up to .9
The current values of the statistics calculation.
$Q sends, e.g.
ActN: Current value of the individual results
"3"
Data for MN1:
Mean: Mean value (decimal places as in result)
"3.421"
Std: Standard deviation (1 decimal place more than in result)
"0.0231"
RelStd: Relative standard deviation (in %, 2 decimal places)
"0.14"
3.2.2.65. Info.SiloCalc.C24.Name read
only
Info.SiloCalc.C24.Value read
only
Info.SiloCalc.C24.Unit read
only
for
.C25 as for .C24
Info.SiloCalc.C26.ActN read
only
Info.SiloCalc.C26.Mean read
only
Info.SiloCalc.C26.Std read
only
Info.SiloCalc.C26.RelStd read
only
for
.C27 as for .C26
The current values from the silo calculations. C26 is the mean value out of the
C24 variables; C27 comes from C25.