6 appendix, 1 technical specifications – Metrohm 709 IC Pump User Manual
Page 57

6.1 Technical specifications
709 IC Pump
49
6 Appendix
6.1 Technical
specifications
Type
Serial dual piston with two valves
Pump capacity
Flow range
0.05...5.00 mL/min
Flow accuracy
< 3.5 % (at 150 bar and 1 mL/min)
Flow constancy
< 0.5 % (after unit warmed up)
Reproducibility of
eluent flow
±
0.1 %
Pressure
Pressure range
2.709.0010:
0...50.0 MPa (0...500
bar)
2.709.0110 (metal-free): 0...30.0 MPa (0...300
bar)
Resolution for pressure
indicator
0.1 MPa (1 bar)
Pressure measurement
Piezoresistive measurement principle
Response time: 3 ms
Measurement volume: ca. 50
µ
L
Accuracy of pressure
measurement
< 3 %
Residual pulsation
2.709.0010:
< 7 %
2.709.0110 (metal-free): < 10 %
(measured at pump head at 2 mL/min and
100 bar)
Recorder output for
pressure indication
1 mV/MPa (1 mV/10 bar)
Time constant switchable: HIGH = 20 ms
LOW = 1 s
Safety shutdown
Function
Automatic shutdown when upper and lower
pressure limits violated
Maximum pressure limit
Adjustable between 0.1...50.0 MPa (1...500 bar)
Response time: 1 pump cycle
Minimum pressure limit
Adjustable between 0.1 ... 50.0 MPa (1...500 bar),
inactive at 0 bar
Response time: 3 pump cycles
The cutoff function first becomes active 2 min
after the pump flow has been switched on with
the
Pump head
Pump head volumes
Main piston:
40
µ
L
Priming piston:
20
µ
L