5.1 RS232 interface

709 IC Pump

43

5.1.8

Data transmission protocol

The RS232 interface of the 709 IC Pump is configured as DTE (Data

Terminal Equipment) with the following technical specifications:

Standard

Data interface in accordance with EIA standard

RS 232C (DIN 66020, page 1)

Baud rate

9600 (300, 600, 1200, 2400, 4800 can be set

with jumpers on main board, see section 4.1.8)

Data Bits

8

Stop Bits

1

XON/XOFF

off

Control characters C

R

DEC 13

HEX 0D

L

F

DEC 10

HEX 0A

Max. line length

80 characters + C

R

L

F

Cable

For interconnections of the 709 IC Pump with

non-Metrohm units, only a shielded data cable

(e.g. Metrohm D.104.0201) may be used. The

cable shielding must be faultlessly earthed at

both units (pay attention to current loops;

always use star-head earthing). Only connec-

tors with adequate shielding may be used (e.g.

Metrohm K.210.0001 with K.210.9004).

Cable length

max. ca. 15 m

As only the baud rate of the RS232 parameters of the IC pump can be

set by hardware, all other RS232 parameters of the third-party device

must be matched to those of the pump.

5.1.9 Handshake

The handshake outputs (DTR, RTS) are set by the 709 IC Pump, the

handshake inputs (CTS, DSR, DCD), on the other hand, are not tested.

In practical operation with third-party devices, this has the following

consequences:

XON/XOFF must be switched off in the third-party device.

With inquiries of the third-party device to the pump, the re-

sponse of the pump must be waited for before the next inquiry

can be started.