Page 28

background image

3 Operation

709 IC Pump

20

3.2 Flow

rate

The flow rate can be set in the range of 0.05

5.00 mL/min with the

pump drive switched on or off as follows:

1 Select "FLOW"

Press key 4 for ca. 3 s until all LEDs (10...13) go out

and the correction factor appears in display 1.

2 Set flow correction

Press <

> key 5 or <

> key 6 until the desired flow correc-

tion appears in display 1.

3 Store flow correction

Release <

> key 5 or <

> key 6: the value shown in display

1 at the time of release will be stored.

4 Return to standby mode

Press