4 Notes – Maintenance – Faults

709 IC Pump

30

If by mistake an inlet valve

69

is installed instead of the outlet valve

70

,

an extreme pressure buildup occurs within the working cylinder, which

is not detected by the pressure transducer and will destroy the piston

seal

68

!

6 Mounting the pump head

The components of the pump head

18 are reassembled as

shown in Fig. 9. The screw

60 and the piston cartridge 64 may

be tightened only by hand. On the other hand, the two valve

screw holders

71 are tightened with a wrench. The pump head is

then remounted on the pump (see section 3.2).

Inlet valveOutlet valve

6.2824.020/ 6.2824.010/

6.2824.090 6.2824.080

74

75

76

77

78

79

80

74

75

76

79

78

77

80

Fig. 11: Components of inlet valve

69 and

outlet valve

70

74 Valve housing

75 Sealing ring (black)

76 Sleeve

77 Sapphire sleeve

The bright side must point towards

the sapphire sphere

78 Sapphire sphere

79 Ceramic holder for sapphire

sphere

The large recess must point in the

direction of the sapphire sphere

80 Seal

The larger opening must point

outwards