3 Operation
709 IC Pump
22
3.4.2
Maximum pressure limit
The maximum shutoff pressure of the pump can be set in steps of
0.1 MPa from 0.1
…
50 MPa with the pump drive switched on or off. Pro-
ceed as follows:
1 Select "P
max
"
Press
max
" LED 12 lights up.
2 Set shutoff pressure
Press <
↓
> key 5 or <
↑
> key 6 until the desired shutoff
pressure (in MPa) appears in display 1.
3 Store shutoff pressure
Release <
↓
> key 5 or <
↑
> key 6: the value shown in display
1 at the time of release will be stored.
The set limit value should lie between 5.0 and 10.0 MPa above the par-
ticular operating pressure or the maximum admissible operating pres-
sure of the column. If the pump exceeds the preset limit value during
operation, the pump drive is switched off within 1 pump cycle and dis-
abled. If the pump has been switched off by the maximum pressure
limit monitoring, the LED 12 flashes, the shutoff pressure can be re-
quested in the display "P
actual
" (see section. 3.7).
Pressing the
longer flashes and the pump is in the standby mode. Pressing the
The 2.709.0110 IC Pump MF (metal-free version) should only be
operated in the pressure range of 0
…
25.0 MPa (0
…
250 bar). With
higher pressure, the 6.2824.100 PEEK pump head can be damaged.
So make sure that the maximum shutoff pressure
P
max
of the pump is
always set to
≤
25.0 MPa.
3.5
Switching the pump drive on and off
Switching on Pressing the
While the pump drive is running, LED 2 above the
Switching off To switch off the pump drive, press the
again. LED 2 goes out.