7 deaerating the pump – Metrohm 709 IC Pump User Manual

Page 26

background image

2 Installation

709 IC Pump

18

2.7

Deaerating the pump

The purge valve 24 is opened by turning anticlockwise and the pump

can now be deaerated by pressing the key 9.
The solvent bottle should be located somewhat above the pump, but at

least at the same level as the pump head 18 (never below the pump).

When the key is pressed the pump operates at the maxi-

mum flow rate (5 mL/min).

For safety reasons the key

9

functions only in the stop

mode of the pump, i.e. the key is disabled when the pump

is running in normal operation.

Should the pump not prime by itself, the 6.2816.020 syringe supplied is

used to siphon off solvent via the outlet capillary 20 with opened purge

valve through the pump head until there are no more bubbles in the

solvent entering the syringe. The pump can then again be purged

briefly by pressing the key 9.
On completion of this procedure, the purge valve 24 is closed by turn-

ing clockwise. The pump is now ready for operation.