4 rs232 test – Metrohm 709 IC Pump User Manual
Page 42
4 Notes – Maintenance – Faults
709 IC Pump
34
4.4.4 RS232
Test
This test is meaningful only if the 709 IC Pump
is used interconnected with other instruments
via RS 232 connection. In addition, a
3.496.8480 Test plug normally used in the re-
pair service is required for this test. However,
this plug can also be purchased by customers
under the above number.
For the sake of completeness, the procedure
is described here.
(If a diagnostic test of the RS232 interface is
not required, continue with section 4.4.5)
1. Power off.
2. Insert 3.496.8480 plug in port "RS232" 37.
3. Sliding switch 36 on rear panel to setting "RS232".
4. Press and hold
2
5. Release
2 . .
2 0 0
The test runs automatically and checks several functions in
succession. If one of these functions is faulty, the display
"2 0 0" does not appear directly but some value is shown
briefly. If, e.g., no test plug is plugged in, the following appears:
2 . .
2 0 8
2 0 0
6. Power off.
7. Remove test plug.
Connections in the
3.496.8480 plug
TxD
RxD
DCD
DTR
DSR
RTS
CTS