5 error messages, 6 rs232 error rectification – Metrohm 709 IC Pump User Manual
Page 49
![background image](/manuals/586635/49/background.png)
5.1 RS232 interface
709 IC Pump
41
5.1.5 Error
messages
Error messages ‘
EXXX
’ are added to the status messages and sepa-
rated from them by the sign "
;
". The following error messages are pos-
sible:
Error
Meaning
Exit/Corrective action
E29
Wrong value or no value allowed.
Send correct value or call up new
object.
E30
Wrong value or no value allowed.
Send correct value or call up new
object.
E31
Wrong trigger, this trigger is not allowed
or carrying-out of action not possible.
Send correct trigger (exception: $D) or
call up new object.
E32
Pump stopped by automatic pressure
shutdown
(P
min
or P
max
violated).
First press
key.
E39
RS232 receive error; overflow of internal
receive buffer (>82 characters).
First press
key.
5.1.6
RS232 error rectification
Problem
Questions for corrective action
No data transmission
occurs and an error
message appears in the
display of the 709 IC
Pump.
⇒
E01: Transmission error.
Is the cable used properly wired and plugged in?
Is the printer switched on and set to ”on-line”?
⇒
E01: Receive error. Are the RS232 data transmission
parameters of the third-party device and the pump the same?
The received characters
are garbled.
⇒
Are the RS232 data transmission parameters of the third-party
device and the pump the same?
⇒
Does the baud rate of the two devices have the same setting?