The instrument sends the preset value of the active reference


Enter preset. Numeric values must always be preceded by a sign.

Refer to the user’s manual of the specific instrument for special applications.

Program samples

Standard Basic

Simplex cable

Serial port opening and parameters

OPEN "COM1:4800,E, 7, 2, PE"

Power supply setting
(RTS=ON, DTR = ON)
&H3FC register addresse (COM2: &H2FC
)

OUT &H3FC,&H0B

Set DTR line OFF
(RTS=ON, DTR = OFF)

OUT &H3FC,&H0A

Data reading

Line input #1,a$

Duplex cable

Serial port opening and parameters

OPEN "COM1:4800, E, 7, 2, PE"

Power supply setting
(RTS=OFF, DTR = ON)
&H3FC register addresse (COM2: &H2FC
)

OUT &H3FC,&H09

Data request ( will be automatic using this
command)

PRINT #1, "?"

Data reading

LINE INPUT #1,a$

Visual Basic

The communication control (MsComm) of VisualBasic must be applied :

Port opening

' Use COM1.
Comm1.CommPort = 1
' 4800 baud, even parity, 7 data, and 2 stop bit.
Comm1.Settings = "4800,E,7,2"
' Open the port.
Comm1.PortOpen = True

P

ower supply setting

' Simplex Cable'

Form1.MSComm1.DTREnable = True
Form1.MSComm1.RTSEnable = True

' Duplex Cable'

Form1.MSComm1.DTREnable = True
Form1.MSComm1.RTSEnable = False

Data request

' Simplex Cable'

MSComm1.DTREnable = False
Timer1.Interval = 150
Timer1.Enabled = True

'Duplex cable + duplex instrument'

MSComm1.Output = "?" + Chr$(13)

'Duplex cable + simplex instrument'

MSComm1.Break = True
'Incremente Timer1.Interval in case of no transmission'
Timer1.Interval = 10
Timer1.Enabled = True
MSComm1.Break = False

Data reading

InString$ = Comm1.Input

For more information, refer to the help menu of MSComm in Visual Basic. Program available on www.sylvac.ch