Fault Location Mode

35

For example, for a cable with a dielectric constant of
2.778, the velocity percentage is 60%. If the cable is
100 feet long, then the Site Analyzer should be set to a
frequency span of 700 MHz. The resolution of the
graph will be 5 inches.

Press the

Mode key

Press from

Select Mode

780.00

MHz

2500.00

Measure Match Mode

Calibration: OFF

99.99

1.00

V

S

W

R

Save/

Recall

Auto

Scale

Limit

Test OFF

HOLD

Print

Measure

Match

Fault

Location

Press the

Config key

Press to

display the

frequency

settings

780.00

MHz

2500.00

Measure Match Mode

Start:

780.00

MHz

Stop: 2500.00 MHz

99.99

1.00

V

S

W

R

Freq

Scale

Limit

Line

Cable

Loss

Center: 1640.00 MHz

Span: 1720.00 MHz

Band List: Custom

Freq

Scale