Cleaning – UVP BioChemi System with EpiChemi II Darkroom User Manual
Page 38

Remove bulbs by rotating the bulbs,
approximately a ¼ turn, and pulling at
the same time, much like replacing
fluorescent overhead bulbs. Then
replace the bulbs in the same
configuration that they were originally
installed; UV bulb, White bulb, and UV
bulb configuration.
Replace the cover and test lighting with
the outside switch on the EPICHEMI II.
*Note: When testing bulbs, be sure
to wear proper UV protecting
eyewear and any other appropriate
personal protective equipment.
Cleaning:
Zoom or fixed lens: Gently wipe the glass surfaces with a lens cloth, do not use
any abrasives or detergents.
Transilluminator: Gently wipe with a non-abrasive cloth to clean. You can use
distilled or deionized water to rinse the surface.
Fold-down platform: Wipe with a non-abrasive cloth to clean. You can use
distilled or deionized water to rinse the surface.
WWW.UVP.COM
US Technical Support: 1.800.452.6788
UK Technical Support: +44.0.1223.420022
38