beautypg.com

UVP BioChemi System with EpiChemi II Darkroom User Manual

Page 24

background image

6. If you want to preview and acquire for the same exposure, the Lock

Exposure Times needs to be checked in the Acquire menu.

7. Place the UVP target or printed text on the transilluminator.
8. Zoom in all the way and focus on the lettering on the gel cutter.
9. Adjust the aperture on the zoom lens until you can see your bands

without saturating the image. Oversaturated areas in the preview
will turn red. Make sure there is no red in the image before capture.
You can click the Minimal Dialog button to minimize the menu
area.

10. Click the Capture button to capture the image (saturation areas in

the preview will turn red).



Capture Button

11. If the image captured is not satisfactory, adjust the time on the

integration tab and repeat steps 9 through 11.