UVP BioChemi System with EpiChemi II Darkroom User Manual
Page 27
No Integration – straight accumulation of signal on the chip. The image
captured is representative of all of the information on the chip at the end of
the capture time period.
Application: Acquiring blots with known exposure times.
1.
With the Options box open, click on the Start Preview button.
2.
If not already selected, click on the Acquire tab (On Chip,
Sequential Integration and Dynamic Integration should not be
selected) and adjust the exposure time to the desired amount of
time.
Acquire Button
Capture
3. Set
desired
Exposure Time in minutes, seconds or millisec.
4. Set
Binning: 2 x 2 (horizontal x vertical) recommended
5. Click
Capture.