3.2 EMC COMPATIBILITY
Sunnex Inc.
Electromagnetic Compatibility User Information for the Celestial Star MRI Light
WARNING: Medical Electrical Equipment needs special precautions regarding EMC and needs to be
installed and put into service according to the Electromagnetic Compatibility [EMC] information provided in
the accompanying documents provided in the Appendix.
WARNING: Portable and Mobile RF Communications Equipment can affect Medical Electrical Equipment.
WARNING: The equipment or system should not be used adjacent to or stacked with other equipment and
that if adjacent or stacked use is necessary, the equipment or system should be observed to verify normal
operation in the configuration in which it will be used.
NOTE: The EMC tables and other guidelines that are included in the Instruction Manual provide information
to the customer or user that is essential in determining the suitability of the Equipment or System for the
Electromagnetic Environment of use, and in managing the Electromagnetic Environment of use to permit the
Equipment or System to perform its intended use without disturbing other Equipment and Systems or non-
medical electrical equipment.
To avoid electromagnetic interference, guidelines in this instructions manual and guidelines of MRI MUST
be strictly followed.
EMC Test Summary
TEST STANDARD
RESULTS
IEC 60601-1-2:2001
w/A1:2004 Example: Basic
Standards from IEC 60601-1-
2:2001 w/A:2004
SUB-TEST TEST
PARAMETER
COMMENT
Radiated and Conducted
Emissions
CISPR 11:2004
Emissions below Group 1, Class B limits
Pass
Harmonic Emissions
IEC 61000-3-2:2001
Emissions within specified Class A limits
Pass
Voltage Fluctuation and
Flicker
IEC 61000-3-3:2002
4% dmax.
Pass
Electrostatic Discharge
IEC 61000-4-2:2004
+/-6kV Contact, +/-8kV Air
Pass
Electrical Fast Transients
IEC 61000-4-4:2004
+/-0.5,1,2kV AC Mains and P/S to Lamp
Pass
Surges
IEC 61000-4-5-2001
+/-1kV Line to Line
+/-2kV Line to Protected Earth
Pass
Voltage Dips and Dropouts
IEC 61000-4-11:2004
5% for ½ Cycle, 40% for 5 Cycles, 70% for 25
Cycles, 5% for 5 Seconds
Pass
(See Notes)
Notes: A UPS should be used if mains quality is suspect.
3.3 TRANSPORTATION AND STORAGE
This light is packaged in card board and should be transported protected from moisture. The light should be
stored in a dry environment and room temperature.