Sunnex CS2050D-MRI (Dual Ceiling) User Manual
Page 17

Sunnex Celestial Star MRI Service & Installation Manual
17
Figure 6: Celestial Star ceiling MRI mounting with horizontal plate and L-bracket (Recommendations)
Figure 7: Celestial Star MRI L-bracket (Recommendations)
NOTES
1. These are recommendations only; Sunnex is not responsible for all the material required for
mounting. Bolts, nuts, horizontal plates, L-brackets and all other materials required for mounting the
light on ceiling supplied by the Contractor.
2. Contractor has the final responsibility for the strength and stability of the Mounting Structure.
Angle 50 deg to 70 deg, to
be determined at site
Side plates for support
This side of L-bracket to be mounted
square, otherwise the light may sway
4cmX4cmX1cm , thk