Sunnex Celestial Star MRI Service & Installation Manual

37




Guidance and Manufacturer’s Declaration Emissions Equipment

and Systems that are NOT Life-Supporting

The Celestial Star MRI Light (CS2050M-MRI, CS2050C-MRI, CS2050ME-
MRI, and CS2050CE-MRI) is intended for use in the electromagnetic
environment specified below. The customer or user of the Celestial Star
MRI Light (CS2050M-MRI, CS2050C-MRI, CS2050ME-MRI, and
CS2050CE-MRI) should ensure that it is used in such an environment.

Immunity

Test

IEC

60601

Test

Level

Complia

nce

Level

Electromagnetic Environment

Guidance

Conducted
RF
IEC 61 000-
4-6

3 Vrms
from

150 kHz
to
80 MHz

V1 = 3
Vrms

Portable and mobile RF
communications equipment should
be separated from the Celestial Star
MRI Light (CS2050M-MRI, CS2050C-
MRI, CS2050ME-MRI, and
CS2050CE-MRI) by no less than the
recommended separation distances
calculated/listed below:
D = (3.5/V1)

P

Radiated
RF
IEC 61000-
4-3

3 V/m

80 MHz
to
2.5 GHz

E1 =
3V/m

D = (3.5/ E1)/

P

80 to 800 MHz

D = (7 / E1)/

P

800 MHz to 2.5 GHz


Where P is the maximum power
rating in watts and D is the
recommended separation distance in
meters.

Field strengths from fixed
transmitters, as determined by an
electromagnetic site survey, should
be less that the compliance levels
(V1 and E1).

Interference may occur in the vicinity
of equipment containing a transmitter.