Recommended Separation Distances Between Portable and
Mobile RF Communications Equipment and the Celestial Star
MRI Light (CS2050M-MRI, CS2050C-MRI, CS2050ME-MRI, and
CS2050CE-MRI) Equipment and Systems that are NOT Life-
Supporting
The Celestial Star MRI Light (CS2050M-MRI, CS2050C-MRI,
CS2050ME-MRI, and CS2050CE-MRI) is intended for use in the
electromagnetic environment in which radiated disturbances are
controlled. The customer or user of the Celestial Star MRI Light
(CS2050M-MRI, CS2050C-MRI, CS2050ME-MRI, and CS2050CE-
MRI) can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF Communications
Equipment and the Celestial Star MRI Light (CS2050M-MRI,
CS2050C-MRI, CS2050ME-MRI, and CS2050CE-MRI) as
recommended below, according to the maximum output power of the
communications equipment.
Recommended Separation Distances for the
Celestial Star MRI Light (CS2050M-MRI,
CS2050C-MRI, CS2050ME-MRI, and CS2050CE-
MRI)
Maximum
Output Power
(Watts)
150 kHz to 80
MHz
d = 1.1667
P
80 MHz to
800MHz
d = 1.1667
P
800 MHz to 2.5
GHz
d = 2.3333
P
0.01 0.11667 0.11667 0.23333
0.1
0.36894 0.36894 0.73785
1
1.1667 1.1667 2.3333
10
3.6894 3.6894 7.3785
100 11.667 11.667 23.333