Sunnex Celestial Star MRI Service & Installation Manual

20












Figure 9. Schematic for option 1 (Installing the transformer box outside MRI room using RF

filter)





Option 2 – Installing the transformer box inside the MRI room.

The light is equipped with 30ft of cord on the secondary side and 10ft of cord on the primary side. If installing
the transformer box inside the MRI room. The total length of cord on the secondary side should not be
altered from the original 30ft.

1. Mount electrical box securely to a structural member in the ceiling (ceiling models only) outside of the

safe perimeter (100 Gauss line/10mT) of MRI machinery with a bolt 1/2" diameter x 2" L. (Four holes
are provided on the base plate of transformer box).

2. It is recommended not to mount the electrical box on wall.
3. An access panel should be constructed to allow access to the transformer box for maintenance and

repair such as fuse replacements.

4. Velcro is provided on the base plate of the electrical box and a rope and snap-on hook is provided on

the electrical box to assist in proper installation.

5. Plug lamp lead connector into the receptacle on the side of the electrical box. CAUTION: Lamp lead

connector has a mechanical latch. Be sure lamp connector is fully engaged into electrical box

Celestial

Star MRI

Power

Supply


RF Filter

MRI ROOM

Celestial Star MRI

Ceiling Light

Celestial Star MRI

Mobile Light

From RF

filter to MRI

Light

Output from

Transformer Box

to RF Filter

Hospital Grade

Cord with Plug

RF Panel