Sunnex Celestial Star MRI Service & Installation Manual

21

receptacle and the latch is fully engaged. (To disconnect, press lever on the receptacle and pull the
lamp head connector.)






Figure 10. Lamp lead connection with electrical box



6. Plug electrical box lead into a hospital grade electrical socket.
7. Lamp is ready for use.

CAUTION! THE ELECTRICAL BOX CONTAINS FERROUS MATERIALS AND MUST BE
HANDLED WITH CAUTION. THE TRANSFORMER BOX SHOULD BE SECURED TO A
STRUCTURAL MEMBER AT A SAFE DISTANCE (100 GAUSS LINE) AWAY FROM THE
MRI MACHINERY. DO NOT MOUNT BOX DIRECTLY TO SHEET ROCK OR PLYWOOD
SUBSTRATES. IDEALLY, THE TRANSFORMER BOX IS SECURED BY NON-FERROUS
FASTENERS, DIRECTLY TO THE FLOOR/CEILING OUTSIDE THE MAXIMUM
RECOMMENDED GAUSS LINE.

Option 3 – Hardwiring the light to mains supply

Instructions to hardwire Celestial Star MRI light
The hardwiring instructions apply to following models only:

1. CS2050C-MRI-HW (Celestial Star ceiling MRI, 120V)
2. CS2050CE-MRI-HW (Celestial Star ceiling MRI, 230V)

CAUTION: A LICENSED ELECTRICIAN MUST BE USED TO
PERFORM HARDWIRING INSTALLATIONS, NATIONAL
ELECTRICAL CODE (NEC) MUST BE FOLLOWED DURING
HARDWIRING INSTALLATION.


The Celestial Star lights may be hardwired with mains supply. During such
hardwiring, following instructions must be used:



Lamp head

connector

Electrical

box