Introduction – Sunnex CS2050D-MRI (Dual Ceiling) User Manual
Page 5

Sunnex Celestial Star MRI Service & Installation Manual
5
2. INTRODUCTION
The objective of this manual is to assist you in assembling your Celestial Star MRI light, to introduce some
of its most important features, and to provide suggestions for its care.
2.1 INTENDED USE
The Sunnex Celestial Star MRI light is intended for use in MRI environments where illumination is required
for medical procedures.
CAUTION! This light is intended for out of bore procedures only. Any
deviation or inappropriate use may result in serious injury.
2.2 SAFETY INSTRUCTIONS
The Sunnex Celestial Star MRI light must be operated with the safety of the user and patient in mind. Please
read this manual and any other instructions before installing the equipment. Please make sure that all
appropriate service staff is informed and educated on the equipment to ensure continued safety and needed
maintenance. Be sure you understand the intended use of this equipment and recommended operation
before attempting to install and use the equipment.
IMPORTANT!
• This light is MRI conditional and can be used as a medical procedure light
around the magnet. NOTE! It is extremely important that all guidelines are
strictly followed to ensure safety for users and patients.
• Use the light only for its intended purpose. It is a medical procedure light
intended for use outside of the bore.
• WARNING! The power supply contains ferrous materials. DO NOT bring the
power supply close to the magnet (100 Gauss line/10 mT). ALWAYS take the
furthest route away from the magnet when moving the power supply in the
MRI room.
• The power supply must be securely fastened to a fixed structure if mounted
inside the MRI room.
• ALWAYS bring the light outside the MRI room when performing any service
on the light. This includes any cleaning and/or bulb replacements. The ceiling
light balance arm can be detached and brought out of the MRI room for
maintenance or repair. Assemble the mobile light outside of the MRI room.
• The safety guidelines of your MRI facility must NEVER be compromised. If in
doubt, consult with your facilities manager on proper procedures.