Sunnex Celestial Star MRI Service & Installation Manual
11
4.1.3 INSTALLATION IN THE MRI ROOM
1. Mount electrical box securely to a structural member outside of the safe perimeter (100 Gauss
line/10mT) of MRI machinery with a fastener 1/2" diameter x 2" L. (Four holes are provided on the
base plate of transformer box).
2. It is recommended not to mount the electrical box on wall.
3. Velcro is provided on the base plate of the electrical box and a rope and snap-on hook is provided on
the electrical box to assist in proper installation.
4. Plug male connector (from lamp cord) into the female connector on the electrical box. CAUTION:
Male connector has a mechanical latch. Be sure lamp connector is fully engaged into electrical box
female connector and the latch is fully engaged. (To disconnect, press tab on the female connector
[and hold] and pull the male connector.)
Figure 2. Lamp connector connection with electrical box
CAUTION
1. IN CASE YOU NEED TO DETACH MALE CONNECTOR FROM FEMALE CONNECTOR OF
POWER SUPPLY, YOU MUST FIRST “PUSH” ON THE TAB OF FEMALE CONNECTOR AND
THEN PULL THE MALE CONNECTOR OUT.
2. PULLING THE MALE CONNECTOR FIRST, WITHOUT “PUSHING” THE TAB ON FEMALE
CONNECTOR MAY MAKE THE CONNECTION TIGHT.
3. IF THE CONNECTION FEELS TIGHT, [1] PRESS ON THE MALE CONNECTOR FIRST [2]
“PUSH” ON THE TAB OF FEMALE CONNECTOR AND HOLD [3] PULL THE MALE
CONNECTOR OUT.
5. Plug electrical box lead into a hospital grade electrical socket.
6. Lamp is ready for use.
Lamp Male
connector
Electrical box
Female
connector