background image


Sunnex Celestial Star MRI Service & Installation Manual

14

4.2.2 MOUNTING WITH AN ANGLE L-BRACKET AS SUPPORT

CAUTION: WHEN MOUNTING CELESTIAL STAR CEILING MRI LIGHT, CARE MUST BE
TAKEN TO USE ONLY NON-MAGNETIC MOUNTING STRUCTURE AND HARDWARE.



An angle L-bracket can be used to secure the mounting bracket to a vertical plane above the drop (false)
ceiling. Please refer to the dimensions below to determine the appropriate distance to the drop ceiling based
on the desired clearance under the lowest point of the extension arm. The bottom of mounting bracket
MUST be mounted at 6.5” from the drop ceiling to achieve 75” (plus or minus 1”) from the light to floor for a
floor to drop ceiling distance of 8’.

Figure 4: Mounting with an angle L-bracket (Celestial Star MRI Ceiling)

It is to be noted that 6’ 3” (1,9 m) is the recommended distance from the bottom of extension arm to floor. A
lower installation height may affect head clearance for tall people and a higher installation height may affect
the reach of the light and the user. However, it is up to the user to determine the most appropriate mounting
height. The following table can be used to determine where to mount the transformer bracket relative to the
drop ceiling. The recommended minimum is 2”. The distance ‘B’ in Table 1 should be chosen depending on
the requirements of the end user. The distance between drop ceiling and solid ceiling should also be
considered. If no drop ceiling is present, please refer to Figure 4 for the appropriate mounting height.

This manual is related to the following products: