2 handle sterilization, 3 bulb replacement – Sunnex CS2050D-MRI (Dual Ceiling) User Manual
Page 29

Sunnex Celestial Star MRI Service & Installation Manual
29
7.2 HANDLE STERILIZATION
The Delrin handles may be sterilized as you would any of your stainless steel instruments. However, it is of
sufficient diameter to engage a disposable sleeve.
The Delrin handle is an accessory for use with the Celestial Star (CS2050C/M-MRI) Specialty light. It is
recommended that it is treated as a stainless steel item when submitting it to sterilization process.
Two options are recommended:
1. A flash cycle at 275 – 279
O
F (135 – 137
O
C), for 16 minutes
2. A standard cycle at 275 – 279
O
F (135 – 137
O
C), for 4 minutes
(Above flash cycle and standard cycle are according to sterilization method in AAMI TIR No. 12-1994:
Designing, Testing, and Labeling Reusable Medical Devices for Reprocessing in Health Care Facilities: A
Guide for Device Manufacturers.)
CAUTION ! ANY DEVIATION FROM THESE GUIDELINES MAY RESULT IN DAMAGE
OF THE PRODUCT AND YOUR EQUIPMENT.
7.3 BULB REPLACEMENT
Only use the following bulbs:
Sunnex part# 18560
3x35W: Sylvania Tru-Aim Titan FRB 35MR16Q/10/NSP/T
3x35W: Osram IRC 35W 10°
Bulb replacement - 1
Bulb replacement - 2
Figure 8
Figure 9
CAUTION: ONLY USE OSRAM/SYLVANIA TRU-AIM TITAN BULBS, ANY DEVIATION FROM THE
DIRECTIONS MAY SERIOUSLY DAMAGE YOUR LIGHT. 35W IS MAXIMUM ALLOWED.
When a lamp/reflector burns out, ALLOW SUFFICIENT TIME FOR IT TO COOL BEFORE ATTEMPTING A
REPLACEMENT. Once cool, you should:
i.
Ensure the power to the lamp is off.
ii.
Refer to figure 8. Turn the lamp head over so the three glass assemblies face upward.
iii.
With one hand supporting the backside of the lamp, grasp the notched ring assembly (1) and twist it