Sunnex Celestial Star MRI Service & Installation Manual

22


1. The light is supplied with a hard service cord (STOW 18/3, 600V, -35

O

C to

105

O

C, 10 Amp per conductor @ 25

O

C, approved as water resistant, oil

resistant, outdoor use) suitable for ceiling hardwiring applications. This cord
has 3 conductors (Hot, neutral and ground). The earth and neutral
conductors are labeled as follows:

EARTH

NEUTRAL N


This STOW cord runs through a hole in the transformer box. The outer
diameter of the STOW cord is 0.378” (9.66mm). The hole in transformer box
is 0.90” (22.86mm) for a metal conduit of (Trade size ½ or metric designator
16). The conduit must be fixed to the transformer box with use of nut.

2. Connect the earth conductor with earthing of the facility, neutral conductor
with neutral of mains supply and hot conductor with hot terminal of mains
supply.

3. Use proper wire nuts of appropriate electrical rating considering the STOW
cord specifications and electrical rating of the light.