Starts the Pumping Program operation.
If the Pumping Program was paused, then the Pumping Program resumes at the point where it
was stopped. Otherwise, the Pumping Program starts from Phase 1.
If a Phase number is specified (
specified Phase number.
The pump cannot be started if an alarm condition exists.
E [
Trigger a Pumping Program Event.
Triggers a pre-defined event defined with the Pumping Program’s Event function, causing an
immediate jump to the Pumping Program Phase defined by the event function.
If
specified by
STOP PUMPING PROGRAM
STP
If the Pumping Program is operating, the pump will be stopped and the Pumping Program will be
paused.
If the Pumping Program is paused, the stop command will cancel the pause and reset the
Pumping Program to Phase 1.
VOLUME DISPENSED
DIS
Queries volume dispense only. Set not applicable.
Response:
I
Where: “I
withdrawn.
CLEAR VOLUME DISPENSED
CLD { INF | WDR }
Sets the Dispensed or withdrawn volume dispensed to 0. Command is only valid while the
Pumping Program is not operating. Query is not applicable.
INF
= Dispense volume
WDR
= Withdrawn volume
Query is not applicable.
12.4.3
Configuration and Setup Commands
New settings for any of the following commands will be stored in the non-volatile memory.
DISPENSE PAUSE MODE
PSM [nn | n.n]
Set/query “Dispense, Pause Dispense” mode, where ‘nn’ is the pause time in seconds. If set to 0,
Dispense Pause Mode is disabled
SLOW DOWN MODE
SLO [
Set/query dispense slow down mode. Set enables or disables slow down mode.