3 example 3: ramping the flow rate – New Era NE-9000 Series Programmable Peristaltic User Manual
Page 39

New Era Pump Systems Inc.
www.SyringePump.com
Model NE-9000
Publication #1200-04
31
Phase
Function
5
PS:90
Phase
Function
6
LP:03
Phase
Function
7
BEEP
Phase
Function
8
PS:30
Phase
Function
Rate
Volume
Direction
9
RATE
750 ml/min
2.25 ml
Dispense
Phase
Function
Rate
Volume
Direction
10
RATE
750 ml/min
0.25 ml
Withdraw
Phase
Function
11
LP:EN
11.4.3
Example 3: Ramping the Flow Rate
m
l/
h
r
R
a
te
Time
0
150
200
250
Phases
2 - 4
Phases
5 - 8
Phases
9-11
Note: Graph is representative only.
Pumping rate increments and decrements in steps.
Continuously ramp up and down the pumping rate. Starting at 200 ml/min, the pumping rate will
increment to 250 ml/min in 1.0 ml/min steps after every 0.1 ml has been dispensed. Then the pumping
rate will decrement to 150 ml/min in 1.0 ml/min steps after every 0.1 ml has been dispensed. Finally, the
pumping rate is incremented back to 200 ml/min in 1.0 ml/min steps after every 0.1 ml has been
dispensed, then the process is repeated.
Phase
Function
Rate
Volume
Direction
1
RATE
200 ml/min 0.1 ml
Dispense
Phase
Function
2
LP:ST
Phase
Function
Rate
Volume
Direction
3
INCR
1.0
0.1 ml
Dispense
Phase
Function
4
LP:50
Phase
Function
5
LP:ST
Phase
Function
Rate
Volume
Direction
6
DECR
1.0
0.1 ml
Dispense
Phase
Function
7
LP:99