N.nn, Special setup key, Learn key – New Era NE-9000 Series Programmable Peristaltic User Manual
Page 22: Special power-up functions

New Era Pump Systems Inc.
www.SyringePump.com
Model NE-9000
Publication #1200-04
14
The display will consecutively display, for about 2 seconds, each Setup Configuration parameter and its
current setting. Pressing any non-arrow key will immediately advance to the next Setup Configuration
parameter.
To change a Setup Configuration parameter, press an arrow key under the parameter’s value. To store the
new value, press any non-arrow key or wait 2 seconds. If the parameter value differs from its previous
value, the display will blink. The new parameter value will be stored and the next parameter will be
displayed. See section 10, “Setup Configuration” for a complete description of the Setup Configurations.
After the last configuration parameter is displayed, the display reverts back to displaying the peristaltic
tube diameter. Any new parameter value will take effect immediately upon being stored.
6.15
Special Setup Key
While turning on power to the pump, press and hold the ‘Setup’ key to access the special setup menu. The
first menu entry “Set Tube Diameter” will be displayed:
6.16
Learn Key
While turning on power to the pump, press and hold the ‘Learn’ key to access the Learn mode menu. The
first menu entry “Learn and Repeat” will be displayed
6.17
Special Power-Up Functions
The following special functions are accessed by pressing the relevant key, while turning on power to the
pump.
6.17.1
Firmware Version Display
To display the pump’s firmware version, press the left-most up arrow key (
/ ) while turning on
power to the pump. The display will show:
n.nn
, where ‘n.nn’ is the firmware version number.
Pressing any key will clear the display.
6.17.2
Reset the Pump
To reset the pump, press the right-most up arrow key (
) while turning on power to the pump. The
display will show
. Pressing any key will clear the display.
With a pump having as many complex features as the NE-9000, it is easy for a novice user experimenting
with the pump's setup to get the pump into a 'weird' state. Performing this reset function will bring the
pump out of a 'weird' state.
6.17.3
Learn and Calibrate Selection
Pressing the 'Rate'/'LEARN' key while turning on power to the pump will display the Learn selection
menu. Use the up arrow keys (
) to scroll through the selections. Press a non-arrow key or wait for the
time out to select.
Learn and Repeat Dispense
Tubing calibration
Exit without making any selection.