Loading peristaltic tubing, 1 remove tubing from pump – New Era NE-9000 Series Programmable Peristaltic User Manual
Page 14

New Era Pump Systems Inc.
www.SyringePump.com
Model NE-9000
Publication #1200-04
6
♦ Finally, the peristaltic tube can be loaded and the pump started.
4.
Loading Peristaltic Tubing
! Remove tubing from pump when not in use. Tubing will become
deformed changing dispensed volume.
4.1 Remove Tubing from Pump
4.1.1. Remove liquid from tubing.
1) Turn on power to the Pump.
2) Press and hold Start/Stop and purge out fluid in Tubing.
3) Remove tubing from liquid.
4) Turn off power to the pump.
!
Warning: Liquid will siphon through tubing when it is removed from pump.
!
Warning: Dangerous Liquid in the tubing may cause injury.
4.1.2. Uninstall Cassette
Turn Cassette counterclockwise until it stops. Then pull it out.
4.1.3. Remove Tubing
1. Remove one side of peristaltic tubing from Cassette.
2. Pull out Tubing. Hold rotor from falling.
3. Remove the other side of tube from Cassette.