Assign precrack values, Calculator – MTS Fracture Template User Manual
Page 80
Assign Precrack Values
Specify these parameters if the specimen was precracked on another load frame. When the precrack
parameters are specified here, you can then skip Precrack Specimen in the Main Menu window.
Precrack Parameters
Description
Parameter
Specify the number of precrack cycles
completed.
Precrack Cycles
Completed
Specify the final crack size calculated from
compliance.
Precrack Final Crack Size
Specify the final load during precrack.
Precrack Final P
Specify the maximum load measured during
precrack.
Precrack P Maximum
Specify the final K during precrack.
Precrack Final K
Specify the maximum K value during
precrack.
Precrack K Maximum
Specify any additional comments.
Precrack Comments
Calculator
Use the Calculator to calculate the load to use for new specimens. You can also use the Calculator to
calculate stress intensity (K). Load is calculated from the crack size value and stress intensity (K) value.
Stress intensity (K) is calculated from the crack size value and load value.
Typically, you use the Calculator after the crack size is checked. If the crack size check is performed, the
Multipurpose Elite application supplies the crack size value for the Calculator. You can also enter a crack
size value manually. You must enter the load or stress intensity value manually.
Calculate Load P
Load P Parameters
Description
Parameter
Specify the crack size value to use in the load
calculation.
Crack Size
80 | MTS TestSuite
Run the CTOD Fracture Toughness Test