Page 24

background image

Warning

:

Actuator movement can occur when you install a specimen.

Unexpected actuator movement can result in personal injury, as your hands will be in a
crush zone during this step.

Ensure that hydraulic power is turned off before placing your hands near the actuator. Read
and understand the safety information in the Safety manual, Series 793 Software manuals,
and Load Frame product manuals.

2. Install the specimen in the grips.

3. Zero the control channel sensor output signal.

4. Install the specimen gage.

5. Zero the gage sensor output signal.

Disable Manual Control

To disable manual control of the Station Manager controller:

1. In the Station Manager application, click Manual Command button in the Station Controls panel. The

Manual Command window opens.

2. Deselect the Enable Manual Command check box and close the window.

3. Turn off hydraulics in the Station Manager application.

a) Click HSM Power Off.
b) Click HPU Power Off.

24 | MTS TestSuite

Test Setup