Memmert HCP 246 Humidity Chambers (Generation 2003) User Manual
Page 44

HCP
page 44
Supply failure in operating mode “Normal operation“
After a supply failure the operation is continued with the set parameters. The instant and duration of the
supply failure are documented in the record memory.
Supply failure in operating mode “Weekly programmer“
After a supply failure the operation is continued with the set parameters. The instant and duration of the
supply failure are documented in the record memory.
Supply failure in programme operation
After a supply failure lasting less than 15 minutes the current programme is continued at the point where
it was interrupted. The instant and duration of the supply failure are documented in the report memory.
On a supply failure lasting longer than 15 minutes the oven immediately starts in manual operation for
safety reasons and all settings are set to safe default values (see table).
Supply failure in remote operation
On a supply failure in remote operation the oven immediately starts in manual operation for safety reasons
and all settings are set to safe default values (see table). Programme continuation has to take place from
the PC. The instant and duration of the supply failure are documented in the report memory.
Parameter
Default-Werte
Temperature 20 °C
rh
off