Memmert HCP 246 Humidity Chambers (Generation 2003) User Manual
Page 23

HCP
page 23
14. Select temperature of third ramp segment
Turn the push/turn control clockwise until the temperature display is
flashing.
Hold down the SET key and set the required temperature setpoint of
37.0
°C using the push/turn control.
°C
15. Select relative humidity of third ramp segment
Turn the push/turn control clockwise until the humidity display is
flashing.
Hold down the SET key and set the required humidity setpoint of 50.0
%rh using the push/turn control.
%
rh
16. Set closure command for third ramp segment
Turn the push/turn control clockwise until a segment closure command
(e.g end ) appears.
Press the SET key briefly to enter.
17. Exit programme writing mode EDIT
Turn the push/turn control clockwise until EXIT appears on the display.
Press the SET key briefly to enter.
18. Set temperature monitor
Turn the push/turn control clockwise and set the temperature monitor.
(see section 12)
MIN
AUTO
MAX
°C
19. Start programme
Turn the push/turn control anticlockwise until the stop symbol
is
flashing.
Hold down the SET key and select Start
with the push/turn control.