Memmert HCP 246 Humidity Chambers (Generation 2003) User Manual
Page 22

HCP
page 22
6. Select temperature of first ramp segment
Turn the push/turn control clockwise until the temperature display is
flashing.
Hold down the SET key and set the required temperature setpoint of
50.0
°C using the push/turn control.
°C
7. Select relative humidity of first ramp segment
Turn the push/turn control clockwise until the humidity display is
flashing.
Hold down the SET key and set the required humidity setpoint of 95.0
%rh using the push/turn control.
%
rh
8. Set closure command of first ramp segment
Turn the push/turn control clockwise until a segment closure command
(e.g. end ) appears.
Hold down the SET key and set SPWT [TH] with the push/turn control.
9. Select duration of second ramp segment
Using the push/turn control select the time indication.
Hold down the SET key and set the time 00.45 using the push/turn
control.
h
10. Select temperature of second ramp segment
Turn the push/turn control clockwise until the temperature display is
flashing.
Hold down the SET key and set the required temperature setpoint of
50.0
°C using the push/turn control.
°C
11. Select relative humidity of second ramp segment
Turn the push/turn control clockwise until the humidity display is
flashing.
Hold down the SET key and set the required humidity setpoint of 95.0
%rh using the push/turn control.
%
rh
12. Set closure command of second ramp segment
Turn the push/turn control clockwise until a segment closure command
(e.g. end ) appears.
Hold down the SET key and set nexT with the push/turn control.
13. Select duration of third ramp segment
Using the push/turn control select the time indication.
Hold down the SET key and set the time 00.01 using the push/turn
control.
h