14 Audible alert signals

The interval sound indicates a fault in the temperature control system or the humidifying system. It is
triggered in the following cases:

Fault in the temperature control system:

Tb activ - temperature limiter activated

if the temperature limiter operates

(see section 12.1)

hi-alarm - overtemperature alarm activated

if the overtemperature

protection operates (see section 12.2.3)

lo-alarm- undertemperature alarm activated

if the undertemperature

protection operates (see section 12.2.3)

asf-alarm - temperature outside tolerance

if the automatic monitoring

function operates (see section 12.2.4)

Fault in the humidifying system:

when the water container is empty.

when the humidity exceeds the selected setpoint for longer than 30
minutes.

DOOR OPEN

when the door of the unit is opened for longer than 3 minutes.

The audible alarm can be switched off temporarily by pressing the SET key.